(English) IFM-SM6000

  • Description
  • Reviews (0)

Description

Application
Application totaliser function; for industrial applications
Installation connection to pipe by means of an adapter
Media conductive liquids; water; hydrous media
Note on media
conductivity: ≥ 20 µS/cm
viscosity: < 70 mm²/s (40 °C)
Medium temperature [°C] -10…70
Pressure rating [bar] 16
MAWP (for applications according to CRN) [bar] 17.7
Electrical data
Operating voltage [V] 18…30 DC; (according to EN 50178 SELV/PELV)
Current consumption [mA] 95; (24 V)
Min. insulation resistance [MΩ] 100; (500 V DC)
Protection class III
Reverse polarity protection yes
Power-on delay time [s] 5
Inputs / outputs
Number of inputs and outputs Number of digital outputs: 2; Number of analogue outputs: 1
Inputs
Inputs counter reset
Outputs
Total number of outputs 2
Output signal switching signal; analogue signal; pulse signal; IO-Link; (configurable)
Electrical design PNP/NPN
Number of digital outputs 2
Output function normally open / normally closed; (parameterisable)
Max. voltage drop switching output DC [V] 2
Permanent current rating of switching output DC [mA] 200
Number of analogue outputs 1
Analogue current output [mA] 4…20; (scalable)
Max. load [Ω] 500
Analogue voltage output [V] 0…10; (scalable)
Min. load resistance [Ω] 2000
Pulse output flow rate meter
Short-circuit protection yes
Type of short-circuit protection pulsed
Overload protection yes
Measuring/setting range
Measuring range
0.1…25 l/min 0.005…1.5 m³/h
Display range
-30…30 l/min -1.8…1.8 m³/h
Resolution
0.02 l/min 0.002 m³/h
Set point SP
0.25…25 l/min 0.015…1.5 m³/h
Reset point rP
0.1…24.9 l/min 0.005…1.495 m³/h
Analogue start point ASP
0…20 l/min 0…1.2 m³/h
Analogue end point AEP
5…25 l/min 0.3…1.5 m³/h
In steps of
0.02 l/min 0.002 m³/h
Volumetric flow quantity monitoring
Pulse value 0.00001…30 000 m³
Pulse length [s] 0,01…2
Temperature monitoring
Measuring range [°C] -20…80
Resolution [°C] 0.2
Set point SP [°C] -19.2…80
Reset point rP [°C] -19.6…79.6
Analogue start point [°C] -20…60
Analogue end point [°C] 0…80
In steps of [°C] 0.2
Accuracy / deviations
Flow monitoring
Accuracy (in the measuring range) ± (0,8 % MW + 0,5 % MEW)
Repeatability ± 0,2% MEW
Temperature monitoring
Accuracy [K] ± 2,5 (Q > 1 l/min)
Response times
Flow monitoring
Response time [s] 0.15; (dAP = 0)
Delay time programmable dS, dr [s] 0…50
Damping for the switching output dAP [s] 0…5
Temperature monitoring
Dynamic response T05 / T09 [s] T09 = 20 (Q > 1 l/min)
Software / programming
Parameter setting options Flow monitoring; quantity meter; Preset counter; Temperature monitoring; hysteresis / window; normally open / normally closed; switching logic; current/voltage/pulse output; start-up delay; display can be deactivated; Display unit
Interfaces
Communication interface IO-Link
Transmission type COM2 (38,4 kBaud)
IO-Link revision 1.1
SDCI standard IEC 61131-9
IO-Link device ID 569 / 00 02 39 h
Profiles Smart Sensor: Process Data Variable; Device Identification, Device Diagnosis
SIO mode yes
Required master port type A
Process data analogue 3
Process data binary 2
Min. process cycle time [ms] 5
Operating conditions
Ambient temperature [°C] -10…60
Storage temperature [°C] -25…80
Protection IP 67
Tests / approvals
EMC
DIN EN 60947-5-9
Shock resistance
DIN IEC 68-2-27 20 g (11 ms)
Vibration resistance
DIN IEC 68-2-6 5 g (10…2000 Hz)
MTTF [years] 162
Pressure Equipment Directive Sound engineering practice; can be used for group 2 fluids; group 1 fluids on request
Mechanical data
Weight [g] 542.5
Materials stainless steel (1.4404 / 316L); PBT-GF20; PC; FKM; TPE
Materials (wetted parts) stainless steel (1.4404 / 316L); PEEK; FKM
Process connection threaded connection G 1/2 DN15 flat seal
Displays / operating elements
Display
Display unit 6 x LED, green (l/min, m³/h, l, m³, 10³, °C)
switching status 2 x LED, yellow
measured values alphanumeric display, 4-digit
programming alphanumeric display, 4-digit
Remarks
Remarks
MW = measured value
MEW = Final value of the measuring range
Pack quantity 1 pcs.
Electrical connection
Connection Connector: 1 x M12; Contacts: gold-plated
Diagrams and graphs
Pressure loss

dP Pressure loss
Q volumetric flow quantity

Reviews

There are no reviews yet.


Be the first to review “(English) IFM-SM6000”